Response of Eight Knee Orthoses to
Valgus, Varus and Axial Rotation
Loads
Thomas R. Lunsford, M.S.E., C.O.
Brenda Rae Lunsford, M.S.
Jack Greenfield, B.A., C.O.
Sharron E. Ross, B.S.
Abstract
Motion restriction of eight knee orthoses
were compared for the pathological conditions of valgus, varus and axial rotation. The
eight knee orthoses included the Polyaction,
Lerman, Lenox Hill, ECKO, DonJoy Analog, Pro Am and CTi. Valgus, varus and
axial rotation torques were applied to the
test apparatus and corresponding angular
deformities were measured.
Analog and CTi provided the most valgus
control, Analog the varus control and Lerman and CTi the most axial rotational control. The clinical efficacy of the knee orthoses evaluated depended upon the magnitude of the imposed load and the quality of
fit.
IntroductionThe Normal Knee
The knee is a modified hinge joint capable
of the fundamental motions of flexion and
extension in addition to some rotation and
gliding motion. Control of knee motion is
shared by the capsular structures, intra- and
extra-articular ligaments, the joint contours
and the muscles with their tendons which
cross the joint anteriorly and posteriorly.1,2
While sagittal plane motion (flexion/extension) consists of a large, free arc of approximately 150 degrees, motion in the coronal
and transverse planes are very small and limited by ligamentous and capsular structures.1,2 In a normal and fully extended
knee, axial rotation is less than 20 degrees
and combined varus/valgus is less the 5 degrees. 1,3,4 As the knee is flexed towards 90
degrees, increased axial rotation to 90 degrees may be attained,1,5 while combined
varus/valgus is less than 15 degrees.6 While
the knee is most stable when fully extended,
it is also in this position that it is most vulnerable to injury, especially from lateral/medial
forces.
The Injured Knee
Injury to the knee may exist as fracture of
bone or tearing or rupturing of ligaments and
menisci. Understanding the role of the bony
and soft tissue structure of the knee is mandatory for proper management of the posttraumatized knee.
Specific function of the knee ligaments
and the consequence of their injury have
been described 7,8,9as well as tests to determine the extent of each.10,11 Knee joint instability, secondary to tearing or rupturing of
the soft tissue structures, is clinically described by the resulting excessive motion.
Varus is an unstable knee in which the tibial
plateau leans laterally ("bow-legs") and is
the result of disruption of the lateral structures (mid-third capsular ligament, fibulo-collateral ligament, biceps tendon, and ilio-tibial tract, with or without abnormality of
the medial meniscus).8,9 In valgus, the tibial
plateau leans medially ("knock-knees") and
results from disruption of the medial
structures (medial capsular ligament, posterior/anterior cruciates).8,9 Axial rotation refers to the tibial plateau turning externally or
internally with respect to the femur. Rotation injuries can result from almost any combination of structural damage but primarily
result from anterior cruciate, medial capsular and tibial collateral ligament damage.7,9
Drawer sign refers to the tibial plateau sliding anteriorly or posteriorly with respect to
the femur while the patient's knee is flexed.
The drawer sign is most commonly linked to
tear of either the anterior and/or posterior
cruciate ligaments. However, some feel that
to get a significant anterior drawer sign the
medial capsular ligament must also be damaged.9,11 Lastly, recurvatum is the increased
posterior angle of the leg while the patient's
knee is hyper-extended. Posterior support of
the knee has been found to be offered primarily by the posterior cruciate ligament,
posterior capsule and tibial collateral ligament.7,9
Treatment
The excessive motions allowed by an
injury to the knee need to be controlled via
conservative management to enable healing,
or to protect the knee following surgical repair. Knee orthoses have been used as an
adjunct to the management of the post-traumatized knee and designed to immobilize
after surgery or injury, and/or to protect
joints that were painful due to trauma or
disease.12 Lately, knee orthoses have been
designed to stabilize antero-lateral and antero-medial rotatory motion ,13 medio-lateral
motion,14,15 and excessive flexion/hyper-extension motions.16 There also have been efforts toward preventing pain by providing
positioning support of the patella via an orthosis. 17,18
Designs have been as simple as those reported by Palumbo17 and Levine18 which
consist of straps which surround or support
the patella to those which are customized
with specific biomechanical objectives for individual patients.13'14'15
Evaluation of Orthoses
Although the literature is abundant with
information on the anatomy and pathology
of the knee, few studies are reported that
objectively compare treatment using an
orthosis versus no orthosis and even fewer
compare knee orthoses. A survey was conducted to determine the outcome of collegiate football players wearing versus not
wearing prophylactic knee orthoses.19 Surprisingly this survey showed a greater incidence of injury in those who wore an orthosis than those who did not, regardless of the
type of knee orthosis worn. Butler, et al.20
described 13 knee orthoses commonly prescribed for patients with ligamentous laxity
secondary to rheumatoid arthritis or injury.
Butler commented on the "potential function" of these knee orthoses and was critical
of the short moment arms, the application of
biomechanical forces on compliant soft tissues and suspension. Butler created a prescription guide by subjectively evaluating
warmth, comfort, protection against blows
and limitation of various deforming motions.
Motion restriction was graded as slight, fair
or good. A prescriber with specific and subjective treatment goals could use Butler's
guide to zero-in on one of the 13 knee orthoses described. Knutzen, et al. ,21 compared
an Ace elastic support (Becton Dickinson
Co., Rochelle Park, NJ) to the Lenox Hill
derotation knee orthosis (Lenox Hill Brace
Shop, 2245 B. 84th St., New York, NY) on
runners for flexion, axial rotation and varus/
valgus using electrogoniometric evaluation
of knee joint kinematics. The Lenox Hill
derotation knee orthosis restrained axial rotation of surgically repaired runners' knees
approximately 2.3 degrees and the Ace support had no effect. Basset evaluated the effectiveness of the Lenox Hill orthosis in controlling anterolateral rotatory instability and
combined anteromedial-anterolateral rotatory instability in 36 patients using the standard clinical laxity tests.13 Patients with isolated anterolateral rotatory instability were
generally improved one grade on the Lachman scale while wearing the Lenox Hill knee
orthosis. However, when there was combined instability, excessive anterolateral rotatory instability was unchanged. Hoffman,
et al.,12 evaluated six commercially available
knee orthoses for their ability to stabilize
ligamentous injuries against valgus, axial rotation and the drawer sign symptoms using
fresh cadaver specimens by measuring bony
displacements as external loads were applied. Of the six knee orthoses evaluated by
Hoffman, the 3D 3-Way knee orthosis (3D
Orthopedic Inc., Dallas, TX) provided the
greatest overall restriction of knee motion.
The American Academy of Orthopedic
Surgeons22 reviewed knee orthoses organized into categories of prophylactic, rehabilitative and functional. Generally very little objective biomechanical or functional
data regarding the orthoses were provided
by the manufacturers. However, a cadaver
study was reported wherein the resistance of
four prophylactic orthoses to valgus loads
was measured. They also reported that there
was no limitation of the drawer sign by any of
the four orthoses evaluated. In the category
functional knee orthoses, Robert Hunter,
M.D., compared the effectiveness of the
Lenox Hill and the CTi knee orthoses in
protection against the drawer sign in 15 subjects with an absent anterior cruciate ligament in one knee. The results showed that
both orthoses held the knee to within normal
limits at 15 pounds of anteriorly directed
force at both 25 degrees and 90 degrees of
knee flexion while neither provided protection at 20 pounds of force.
Purpose
Because performance data are scarce and
technical/clinical guidelines unavailable, it is
difficult for a clinician to objectively select
the appropriate knee orthosis for a given patient/pathology. Therefore, the purpose of
this study is to objectively compare eight
commonly prescribed knee orthoses for their
ability to restrain excessive knee valgus, varus and axial rotation. A follow-up study will
be directed at evaluating the drawer sign and
recurvatum.
Specific Aims
The specific aims of this study are to:
- Develop an apparatus for evaluating
valgus, varus and axial rotation;
- Document the mean (+/- sd) of valgus,
varus and axial rotation deformity for
each of several torque loads for eight
knee orthoses;
- Compare the motion allowed among
the eight knee orthoses at each of the
torque loads;
- Define the relationship between the
torque and angle for each orthosis tested, in each position.
MethodMaterials
To objectively evaluate the eight knee
orthoses, a unique testing apparatus was developed (Figure 1)
. The apparatus consists of
an above-knee prosthesis which had been
modified to allow valgus, varus and axial rotation angles much greater than normal. The
socket portion of the prosthesis was filled
with polyester foam and a mounting pipe was
inserted to hold the prosthesis horizontally.
This pipe was inserted through two tee-joints, which were attached to a wood base.
The prosthetic knee joint was removed and
replaced with a custom-made joint constructed of soft crepe. This custom-made
crepe knee joint required minimum torque
to create valgus, varus and axial rotational
deformities.
To simulate natural skin friction, a rubber
sleeve was stretched over the prosthesis.
This allowed the components and straps of
the various knee orthoses to grip the otherwise smooth surface of the prosthesis.
With the test apparatus in the horizontal
side-lying position, the tibial portion freely
fell into 30 degrees of valgus or varus and 25
degrees of axial rotation (Figure 2
and Figure 3
).
Axial rotation was indicated by two offset
alignment marks on the femoral and tibial
sections of the prosthesis. The femoral portion of the AK prosthesis was immobilized
and the tibial section was able to angulate
medially or laterally independent of the femur when torque was applied. The instrumentation used consisted of a torque wrench
(Model 1502DIN, Consolidated Devices
Inc., City of Industry, CA), tensiometer
(Model DDP-50, John Chatillon & Sons,
Inc., Greensboro, NC), electrogoniometer
(Neuromuscular Engineering Department,
Rancho Los Amigos Medical Center, Downey, CA) and linear scale (Figure 4)
. The
electrogoniometer, which consisted of a
truss assembly connected to a potentiometer
and a digital readout, measured and indicated the angular deformities produced. Custom-made mounting brackets were used to
secure the electrogoniometer for measuring
angular deformity of the test prosthesis (Figure 5)
. The brackets could be arranged to
accommodate valgus, varus and axial rotation angulation measurements.
The eight knee orthoses selected for testing included the Polyaction, Lerman, Lenox
Hill, ECKO, DonJoy, Analog, Pro Am and
CTi (Table 1)
. These orthoses were either
purchased or donated. Although all existing
knee orthoses were not evaluated, this group
is considered to be representative of the
knee orthoses presently available.
Procedure
Preliminary calculations indicated that a
150 pound patient in single limb support with
varying degrees of varus/valgus may require
a knee orthosis to restrain approximately 500
inch-pounds of torque. Therefore, for the
purpose of comparison, the applied valgus
and varus torque was set to range from 0 to
650 inch-pounds, in ten 65 inch-pound increments. Similarly, the applied axial rotation
torque ranged from 0 to 240 inch-pounds, in
60 inch-pound increments. These ranges of
applied torques were selected to represent
practical clinical conditions. The extreme
torques associated with high level professional sports is unknown.
For each orthosis, the torque application
sequence was repeated three times. This
loading sequence was also done when no
orthosis was on the testing apparatus. All
testing took place with the test apparatus in
the horizontal supine position.
Once a knee orthosis was placed on the
test apparatus, axial rotation torque was applied with a torque wrench with an analog
dial indicator, while valgus/varus torque was
applied with a constant lever tensiometer.
The lever arm for the valgus/varus tests was
13 inches from the point of application of the
force to the knee joint axis, thus giving
torque (inch-pounds) force (pounds) x
13 inches. Each angular deformity was digitally indicated on the electrogoniometer.
This procedure was repeated three times for
each deformity for each of the eight knee
orthoses. Between trials, slippage and inelastic alignment were corrected.
For each value of valgus, varus or axial
rotational torque applied, the angle that the
fixture deformed, with or without an orthosis attached, was recorded.
Valgus and varus force was applied with
the tensiometer at a point corresponding to
the apex of the lateral malleolus (Figure 5)
.
As the fixture with the orthosis applied began to deform, the resulting angulation was
displayed on the electrogoniometer digital
readout. After each trial was performed, the
electrogoniometer was reset to zero and the
leg reset to its original position as indicated
by alignment marks on the prosthesis and on
the base board.
Axial rotational torque was applied with a
torque wrench to a nut welded to the distal
margin of the prosthesis (Figure 7)
. The electrogoniometer mounting brackets were rotated 90 degrees from the previous location
to record the angles produced by the deforming knee orthosis. Anterior alignment marks
on the femur and tibia sections of the test
fixture were used to reset the fixture to the
starting point.
Data Analysis
The data were screened for outliers and
summarized. Since only one of each orthosis
was tested, the three replicate measurements were averaged and that value used to
represent the motion obtained for each orthosis at each position and torque load. Analysis of variance was used to test for variation
between the orthoses. This was done separately for each motion at each of the imposed
torques. Regression analyses were also used
to detect significant linear trends between
the variables of torque and angle. As the
applied torques caused the apparatus and
orthosis to deform, it was obvious that some
of the resistance to deformity was caused by
the apparatus. To isolate the resistance to
motion caused by the orthoses alone, the
resistance due to the apparatus was subtracted. This was done by subtracting the torque
required by the apparatus alone at a specific
angular deformity from the opposite torque
required by the apparatus and orthosis being
tested. The procedure was repeated at two
degree increments from 0 to 50 degrees.
The statistical programming package
Crunch (Crunch Software Corp., 5335 College Aye, Oakland, CA 94618) was used to
perform the data analysis, and all testing was
done at a .05 level of significance.
ResultsValgus
The mean (+/- sd) valgus angle of deformity
ranged from a low of 0 degrees for the Polyaction, Lerman, ECKO, Analog and Pro
Am orthoses at 65 inch pounds of torque to a
high of 46.7 degrees (+/- 1.16) for the ECKO
knee orthoses at 650 inch-pounds of torque
(Table 2)
. Post hoc testing using the Sheffe
multiple comparisons test showed the Polyaction, Lerman and Analog knee orthoses
had consistently greater ability to resist deformity, than the other five orthoses,
through 390 inch-pounds of torque. At the
load of 455 inch-pounds the Analog knee
orthosis was joined by the CTi in showing
the most significant resistance to deformity
through the final load of 650 inch-pounds
attaining a valgus angle of 13 degrees (+/- 1),
13 degrees (+/- 0) respectively. The Lennox
Hill and ECKO knee orthoses showed the
least resistance to deformity, where 6 degrees (+/- 1) and 7 degrees (+/- 1) of valgus
motion was noted at 130 inch-pounds of applied torque. At 260 inch-pounds the DonJoy joined these two orthoses and continued
to resist valgus motion significantly less than
the other five knee orthoses through the final
torque load of 650 inch-pounds, attaining
46.7 degrees (+/- 1.16), 42.7 degrees (+/- 1.16)
and 38.7 degrees (+/-1.16) for the ECKO,
Lenox Hill and DonJoy, respectively. The
test prosthesis with no orthosis deformed 49
degrees (+/- 1) into valgus (maximum deformity permissible on the apparatus) at 325
inch-pounds and was significantly more flexible than all of the test orthoses (Table 2)
.
The first load of 65 inch-pounds of torque
was the only load at which all of the knee
orthoses protected the knee to within the
maximum comfortable valgus deformity of
3.4 degrees.4 The Polyaction provided the
best protection by allowing less than 3.4 degrees of valgus through 260 inch-pounds of
torque. The Polyaction, Lerman, Analog
and CTi best restrained valgus to within 8
degrees, through 390 inch-pounds of torque,
the angle which corresponds to internal knee
disruption.3(Table 2)
Significant linear trends were detected for
torque versus angle for each of the orthoses
evaluated, (r .95 to .99). Further regression was applied and the best curve fit to
these data were observed with third order
regression (r .983 to .998); however, this
was not significant (Figure 7)
.
Varus
The Polyaction, Lerman, Analog and Pro
Am knee orthoses showed significantly
greater resistance to varus deformity from 65
to 260 inch-pounds of applied torque; 7 degrees (+/- 0), 9.3 degrees (+/- 2.08), 8.3 degrees (+/- .58), and 9.7 degrees (+/- 1.16), respectively, at 260 inch-pounds (Table 3)
. At
325 inch-pounds the Polyaction yielded 11.3
degrees (+/- 1.53) and the Analog yielded 9
degrees (+/- 0) and both were significantly
more resistant to varus motion than any of
the other orthoses. From 390 to 650 inch-pounds of applied torque the Analog was,
alone, significantly more resistant to varus
deformity than the other seven orthoses, attaining a maximum varus deformity of 15.7
degrees (+/- .58). The ECKO knee orthosis
exhibited the greatest degree of flexibility
being significantly more deformed, 49.7 degrees (+/- .58), at the maximum torque load.
The ECKO, DonJoy, Lenox Hill, and CTi
deformed significantly more than all of the
other orthoses for varus loads of 195 inch-pounds to 325 inch-pounds. The test prosthesis, with no orthosis, deformed 50.7 degrees (+/- 1.53) into varus (maximum deformity permissible on the apparatus), at a load
of 455 inch-pounds, which was significantly
greater than any of the test orthoses.
At the maximum comfortable varus angle,
3.4 degrees,4 the Polyaction provided the best
protection (195 inch-pounds) while the
ECKO, Lenox Hill and DonJoy provided the
best protection by exceeding the 3.4 degrees
maximum comfortable varus range in degrees
after only 65 inch-pounds of applied load. At
the minimum angle at which internal joint disruption occurs (8 degrees3), four orthoses -
Polyaction, Lerman, Analog and Pro Am -
were stiffer than the others holding through
260 inch-pounds of varus torque. The ECKO
provided the least protection by exceeding the
8 degrees after the first torque load.
Again a significant linear trend was observed between the variables of torque and
angle for each of the orthoses (r .979 to
.99). With third order regression applied,
there was an improvement in curve fit
(r .991 to .999) which, again, was not significant (Figure 8).
Axial Rotation
There are two groups of applied rotational
torques: 0 to 60 inch-pounds, to evaluate the
test apparatus with no orthosis; and 30 to 240
inch-pounds used to evaluate the eight knee
orthoses.
The greatest variation between the eight knee orthoses occurred in axial rotation. The only axial rotation load at which all eight orthoses could be compared was 60 inchpounds (Table 4)
. At this axial rotation load, the Polyaction and CTi deformed 2.3 degrees (+/-1.16) and 2.3 degrees (+/-.58), respectively; both of these knee orthoses were significantly more resistant to axial rotation motion than any of the others. The ECKO and DonJoy orthoses allowed significantly more motion, 40.3 degrees (+/- 1.53) and 38.3 degrees (+/- 1.16), respectively, at the first load of 60 inch-pounds, and exceeded the maximum permissible axial rotational deformity at the second rotational torque of 120 inch-pounds. Of the remaining six knee orthoses, the Lennox Hill was significantly more flexible than any of the others, allowing 30.3 degrees (+/-3.51) of axial rotation at 120 inchpounds. At 180 and 240 inch-pounds of applied torque, the CTi and Lerman were significantly more resistant to axial motion, attaining 31 degrees (+/- 1) and 34.7 degrees (+/- 1.53), respectively, at the maximum load.
The maximum comfortable axial rotation
angle of 25 degrees5 was maintained up to
180 inch-pounds by only the Lerman and CTi
knee orthoses. The axial rotation angle at
which internal knee disruption occurs (37 degrees) was maintained up to 240 inch-pounds
of applied axial rotation torque by only the
Lerman and CTi knee orthoses.3
There was significant linear relationship
between the applied torque and resulting angle (r .93 to .99) for all of the orthoses
tested for rotation, which improved as third
order regression was applied (r = .996 to
.999). This change was also, not significant.
Discussion
The Analog and CTi knee orthoses exhibited the overall best resistance to high valgus
loads (Table 2)
. It is interesting to note that
although the Analog and CTi knee orthoses
offer the greatest resistance to valgus at high
loads; the Polyaction, Lerman and Analog
were better able to avoid the 3.4 degrees
maximum comfortable deformity up to 195
inch-pounds of torque (Table 2
and Figure 7
).
The Analog knee orthosis exhibited the
overall best resistance to varus torques (Table 3
and Figure 8
), especially at the higher
varus loads (>325 inch-pounds). As was the
case for valgus, the Polyaction knee orthosis
was best able to avoid the maximum comfortable varus deformity (3.4 degrees) up to
applied loads of 195 inch-pounds.
In general there was a 25 percent greater
resistance to valgus than varus; the reason
for this is not clear.
The CTi and Lerman knee orthoses exhibited the overall best resistance to axial rotation torques (Table 4
and Figure 9
), including
the axial angles corresponding to the maximum comfortable rotation (25 degrees) and
protected completely against the axial angle
at which internal joint disruption occurs (37
degrees). Regression curves of the axial rotation data (Figure 9)
show that the CTi and
Lerman knee orthoses were 25 to 30 percent
better than the other six knee orthoses tested. This study did not support the finding of
Knutzen, et al.21 which showed that the
Lenox Hill restricted axial rotation of the
knee to approximately 7 degrees, during
weight-bearing while walking. While these
two studies are not comparable, the Lenox
Hill exceeded 7 degrees of axial rotation at
the first torque load which is calculated to beless than that experienced during the weight-bearing phase of gait.
Only two of the knee orthoses tested required a plaster impression to be sent to the
manufacturer: the Analog and Lenox Hill.
The CTi knee orthosis required an outline of
the leg be drawn by a company representative. Therefore, these three were considered
custom made and all the others were "off the
shelf." In spite of being custom fabricated,
the Lenox Hill knee orthosis did not fit well,
which may account for its poor overall performance. Conversely, the Pro Am knee
orthosis was not custom fabricated, did not
fit well and performed surprisingly.
In general the more rigid the knee orthosis, the more resistant it was to deforming
forces. Rigidity depended upon design
(CTi), the use of metal sidebars (Analog,
Polyaction and Lerman), and overall length
(leverage). If the knee orthosis fits well, then
good resistance deformity at small loads was
more likely and corrective pressures were
less (not measured).
The best knee orthosis in terms of fit and
cosmesis, the ECKO, did not perform the
best. It remains an obvious problem that the
cosmetically "conscious" knee orthosis may
not necessarily be functionally capable of doing the job for which it was intended; yet the
functionally capable knee orthosis may not
be acceptable to the patient or prescriber.
It is the authors' opinion that before a
given knee orthosis can be objectively recommended for a given patient/pathology,
standards should be devised as to how much
knee angle deformity can be tolerated and
the magnitude of external loads which must
be restrained. It is obvious from Teitz, et
al.19 that the orthoses used by the football
players did not provide adequate protection.
However, to determine if protection by a
knee orthosis is even possible, the specific
loads imposed during the activity of football
need to be known. It may be that the most
restrictive knee orthoses may not be capable
of guarding against high deforming torques.
Therefore, the best knee orthosis may not be
good enough. However, it may be found that
the least restrictive knee orthoses may suffice for a wide range of low torque applications.
Thomas R. Lunsford is Chief Orthotist at Rancho Los Amigos Medical Center, Downey, California, and Clinical Director of the Orthotics and Prosthetics Baccalaureate Program at California State University, Dominguez Hills.
Brenda Rae Lunsford is Biostatistician and Manager of Information Systems, Physical Therapy Department, Rancho Los Amigos Medical Center, Downey, California.
Jack Greenfield is Assistant Chief of the Orthotics Department at Rancho Los Amigos Medical Center, Downey, California.
Sharron E. Ross, currently a resident orthotist at Rancho Los Amigos Medical Center, Downey, California, is a graduate of the University of California, Riverside, and also a graduate of the California State University, Dominguez Hills, where this study was conducted in conjunction with the Orthotics and Prosthetics Baccalaureate Program requirements.
References:
- Rasch, P.J. and R.K. Burke, "The Science of
Human Movement," Kinesiology and Applied
Anatomy, Lea & Febiger, 1978, pp. 285-310.
- Welsh, R.P., "Knee Joint Structure and
Function," Clinical Orthopaedics and Related Research, Vol. 147, 1980, pp. 7-14.
- Seering, W.P., R.L. Piziali, DA. Nagel and
D.J. Schurman, "The Function of the Primary
Ligaments of the Knee in Varus-Valgus and Axial
Rotation," Journal of Biomechanics, Vol.13,
1980, pp. 785-794.
- Markolf, K.L., A. Graff-Radford, and H.C.
Amstutz, "In Vivo Knee Stability," Journal of
Bone and Joint Surgery, 60A:5, 1978, pp. 664674.
- Shoemaker, S.C. and K.L. Markolf, "In Vivo
Rotatory Knee Stability," Journal of Bone and
Joint Surgery, 64A:2, 1982, pp. 208-216.
- Markolf, K.L., J.S. Mensch and H.C. Amstutz, "Stiffness and Laxity of the Knee - The
Contributions of the Supporting Structures,"
Journal of Bone and Joint Surgery, 58A:5, 1976,
pp. 583-593.
- Kennedy, J.C. and P.J. Fowler, "Medial and
Anterior Instability of the Knee," Journal of
Bone and Joint Surgery, 53A:7, 1971, pp. 12571270.
- Hughston, J.C., J.R. Andrews, M.J. Cross
and A. Moschi, "Classification of Knee Ligament
Instabilities ä Part I. The Medial Compartment
and Cruciate Ligaments," Journal of Bone and
Joint Surgery, 58A:2, 1976, pp. 159-186.
- Fowler, P.J. "The Classification and Early
Diagnosis of Knee Joint Instability," Clinical Orthopaedics and Related Research, Vol. 147, 1980,
pp. 15-21.
- Gaiway, H.R. and D.L. MacIntosh, "The
Lateral Pivot Shift: A Symptom and Sign of Anterior Cruciate Ligament Insufficiency," Clinical
Orthopaedics and Related Research, Vol. 147,
1980, pp. 45-50.
- Torg, J.S., W. Conrad, and V. Kalen, "Clinical Diagnosis of Anterior Cruciate Ligament Instability in the Athlete," The American Journal of
Sports Medicine, 4:2, 1976, pp. 84-93.
- Hofmann, A.A., R.W.B. Wyatt, M.H.
Bourne and A.U. Daniels, "Knee Stability in Orthotic Knee Braces," The American Journal of
Sports Medicine, 12:5, 1984, pp. 371-374.
- Bassett, G.S. and B.W. Fleming, "The Lenox
Hill Brace in Anterolateral Rotatory Instability,"
American Journal of Sports Medicine, 11:5, 1983,
pp. 345-348.
- Peizer, E., E.J. Lorenze and M. Dixon, "The
Genucentric Joint Orthosis," Medical Instrumentation, 16:4, 1982, pp. 207-208.
Cassvan, A., K.E. Wunder and D.M. Fultonberg, "Orthotic Management of the Unstable
Knee," Archives of Physical Medicine & Rehabilitation, Vol. 58, 1977, pp. 487-490.
- Hastings, D.E., "The Non-Operative Management of Collateral Ligament Injuries of the
Knee Joint," Clinical Orthopaedics & Related Research, Vol. 147, 1980, pp. 22-28.
- Palumbo, P.M., "Dynamic Patellar Brace: A
New Orthosis in the Management of Patellofemoral Disorders," American Journal of Sports
Medicine, 9:1, 1981, pp. 45-49.
- Levine, J., "A New Brace for Chondromalacia Patella and Kindred Conditions," Journal of
Sports Medicine, 6:3, 1978, pp. 137-140.
- Teitz, C.C., B.K. Hermanson, R.A. Kronmal, and P.H. Diehr, "Evaluation of the Use of
Braces to Prevent Injury to the Knee in Collegiate
Football Players," The Journal of Bone and Joint
Surgery, 69A:1, 1987, pp. 2-9.
- Butler, P.B., G.A. Evans, G.K. Rose and
J.H. Patrick, "A Review of Selected Knee Orthoses," British Journal of Rheumatology, 22, 1983,
pp. 109-120.
- Knutzen, K.M., B.T. Bates and J. Hamill,
"Electrogoniometry of Post-Surgical Knee Bracing in Running," American Journal of Physical
Medicine, 62:4, 1983, pp. 172-181.
- Knee Braces," American Academy of Orthopaedic Surgeons: Summer Report, 1984.
|
|