Effect of a Walking Splint
and Total Contact Casts on Plantar
Forces
Andrew Novick, MA., P.T.
James A. Birke, M.S., P.T.
Sandra L. Graham, M.S., P.T.
Edward Koziatek, P.T.
Introduction
Neuropathic plantar ulcers of the foot
have been successfully managed by a variety
of treatment regimens. One of the earliest
methods to promote healing was the application of plaster of paris (POP) casts (1). These
devices reduce localized pressure to the lesion and immobilize the surrounding joints
and soft tissues while preserving functional
ambulation. The casts have either been conventional padded casts (2,3,4), or, more recently, total contact casts (TCCs) using a
minimal amount of padding over the malleoli, posterior heel, and other bony prominences, with additional padding over the lesion (5,6,7,8,9,10).
In cases where a POP cast is either contraindicated or not accepted by the patient,
alternative types of customized footwear
have proven to be effective in healing plantar
lesions. One such device is the posterior
walking splint (5,6). It provides many of the
same benefits as the POP cast, but can be
easily removed and donned when access to
the lesion is required. The POP-molded double-rocker plaster shoe (MDRPS) (4,8,11,12)
and patellar-tendon-bearing (PTB) orthoses
(13,14) have also been used successfully.
The primary cause of neuropathic plantar
ulcers is the cumulative effect of repetitive
mechanical stress to vulnerable areas of the
insensitive foot, usually over a bony prominence (15,16,17,18,19). This information,
combined with the clinical benefits of customized footwear, prompted gait studies to
quantify the effect of these devices on plantar forces and pressure. Birke et al. demonstrated a reduction in pressure under the 1st
and 3rd metatarsal heads (MTHs) with both
a conventional padded POP cast and a TCC
(20). Pollard similarly reported a reduction
in vertical force under the 1st, 2nd/3rd, 4th
and 5th MTHs and diminished mediolateral
(ML) and anteroposterior (AP) shear stresses within a POP cast (21). Birke and Nawoczenski examined the effects of short leg and
PTB walking orthoses and demonstrated reduced pressures under the 1st, 3rd and 5th
MTHs and the heel (22).
As the design of custom footwear evolves
with the acquisition of new knowledge and
materials, these devices must continue to be
studied from clinical and quantitative perspectives. This study examined the walking
splint (SPLINT), the TCC with two different
terminal devices (the rubber walking heel
[TCC/WH] and the cast boot [TCC/CB]),
and a conventional leather oxford shoe
(SHOE). All of these devices have been
proven to be clinically effective in the healing of plantar lesions, but quantitative data
to substantiate the clinical results of the
walking splint, or to determine which reduces plantar forces most effectively, were
not available. Data with the SHOE were also
collected to provide standard force measures
during gait with which to compare the TCC/
WH, TCC/CB and SPLINT treatment conditions.
The specific purposes of this study were to
- substantiate the clinical effectiveness of
the SPLINT by demonstrating a reduction in force at the primary weight-bearing sites of the foot when compared to
the SHOE,
- determine if there was a significant difference in force measures between the
two TCC conditions (TCC/WH and
TCC/CB) as compared to the SPLINT
and
- determine if there was a significant difference in force measures within the
TCC, between the TCC/WH and TCC/
GB conditions.
Additionally, the TCC/CB and TCC/WH
were compared with the SHOE to substantiate the previous findings of Birke et al. and
Pollard (20,21).
Methods and ProceduresSubjects
Twenty normal subjects, 11 males and
nine females, between the ages of 21 and 63
years, participated in this study. The mean
age for all subjects was 36.5 (+/- 10.4) years,
with the mean age for males and females
being 35.5 (+/-6.9) and 37.7 (+/-13.8) years,
respectively. All subjects were free of clinical symptoms in the lower extremity and presented with normal sensation on the plantar
aspect of the right foot.
Instrumentation
Capacitive pressure transducers 2 mm
thick and 1.5 cms in diameter were calibrated as described by the manufacturer (Hercules Orthoflex Data System, Allegheny Ballistics Lab, Cumberland, Md.). The calibration signal for each transducer was sampled
with a Digital Equipment Corp. (DEC) Analog Data Module (ADM) on a DEC Pro3S0
computer (Digital Equipment Corp., Maynard, Mass.). The calibration data was then
transferred to a DEC VAX 11/750 for computation of calibration values. Subject data
collected during the walking trials were similarly collected on the Pro3SO and analyzed
on the VAX. Subject data were sampled at a
rate of 60 hertz for a total of 8.35 seconds,
the longest interval of time allowed by the
ADM software. A four-channel strip-chart
recorder was used to visualize the data being
sampled by the computer, which were converted to force data by factoring out the surface area of the transducers (Gulton TR
4000A, Gulton Industries, Inc., East Greenwich, R.L.). This was done to eliminate size
of the body part as a variable, since the 1.5cm diameter transducers were placed under
the larger calcaneus and midfoot sites, and
the much smaller metatarsal heads.
Testing Procedures
The four transducers were secured to the
plantar aspect of the right foot at the heel,
midfoot (at the level of the navicular), 4th
MTH, and 2nd MTH to provide a representative sampling of forces along both the
length of the foot and across the entire width
of the forefoot. Each transducer was held
securely in place with double-sided tape between the transducer and the skin, then further secured with Transpore tape overlying
the transducer (3M Co., 3M Center, St.
Paul, Minn.) Cotton stockinette covered the
foot and remained intact for the duration of
the study. A second stockinette, which became incorporated into the plaster devices,
was applied over the first. This allowed the
four transducers and two stockinettes to remain as a constant interface between the foot
and inside of all the devices.
Each subject participated in the following
four treatment conditions:
- plaster walking splint (SPLINT),
- total contact cast with a rubber walking
heel (TCC/WH),
- total contact cast with cast boot (TCC/
CB) and
- leather oxford shoe (SHOE).
The testing order of these devices was randomly assigned to avoid bias.
Fabrication of the Plaster Devices
Construction of the total contact casts and
walking splint was similar to that described
by Birke et al. for clinical use (6), but slightly
modified to ensure a standard interface between the foot, transducers and plaster.
None of the plaster devices incorporated
foam padding over a plantar bony prominence, as would be placed over an open ulcer
clinically to further reduce force and promote healing.
Total Contact Cast (Figure 1
and Figure 2
)
Modifications to the described method include the following:
- Webril (Kendall Co., Boston, Mass.)
was used to enclose the toes, which are
normally covered with Sifoam (Knit
Rite, Kansas City, Mo.), and
- two rolls of Scotchcast fiberglass resin
tape (3M Co., 3M Center, St. Paul,
Minn.) were applied to allow weightbearing following a 20-minute drying
period.
If randomization of the treatment conditions required the TCC/WH to be run first, a
layer of plastic food wrap (Dow Consumer
Products, Inc., Indianapolis, Ind.) was
placed over the plantar surface of the initial
layer of fiberglass tape before the rubber
walking heel was secured. This allowed easy
removal of the heel so that the same total
contact cast could be used for both the TCC/
WH and TCC/CB conditions. The cast boot
used for the TCC/CB condition contained a
neoprene crepe sole incorporating a rocker
design, a cut-back heel and a flat sole under
the middle third of the device (AliMed, Inc.,
90Dedham, Mass.).
Walking Splint (Figure 3)
The walking splint was similarly fabricated
according to Birke et al. with the following
modifications:
- the application of Webril under the toes
replacing Sifoam and
- the use of a stockinette instead of Webril to enclose the leg and foot (6).
Data Collection and Analysis
After application of the first device, the
subject assumed the starting position and began walking. Data collection commenced after the subject had walked approximately 20
feet to allow the subject to assume his/her
normal gait characteristics. Gait data were
collected for the allotted 8.35 seconds and
completed prior to the end of the walking
trial to prevent sampling of data while the
subject was decelerating. The same procedure was repeated for each device. All subjects performed two walking trials in each of
the four treatment conditions.
The calibration and raw subject data were
transferred to the VAX for analysis. The
second of the two data sets, which typically
presented less noise when the output was
viewed from the strip-chart recorder, was
used for analysis for most of the treatment
conditions. (This difference was due to the
subject's being more acclimated to walking
in the device after the first trial.) However,
in those few cases when more noise was observed in the 2nd trial due to a misstep by the
subject during gait or to equipment malfunction, the initial trial was used. Transducer
malfunction occurred at one site for two subjects and at two sites for one subject, resulting in the elimination of data sets at these
sites for all four treatment conditions for
these subjects (a total of 16 out of a possible
640 datasets).
A calibration value for each transducer
was initially calculated prior to data collection. The peak force values for all stance
phases contained within the total 8.35-second dataset for each treatment condition
trial were then identified. The first and last
peak values were disregarded, and a mean
peak value for each transducer was calculated for each treatment condition trial for each
subject. The mean peak values for the
SPLINT, TCC/WH and TCC/CB conditions
were also compared to that of the standard
leather oxford shoe to calculate a percent
change in force from the SHOE. An analysis
of variance (ANOVA), SAS, Proc GLM,
was then performed on the force values.
Results
Means and standard deviations for the
force measures are reported in Table 1
. The
ANOVA results (Table 2)
show significant
differences in the force measures among the
four treatment conditions at all four sites
each having a significance level of p<0.001
A Duncan's multiple range test was performed to determine which treatment condition(s) differed (Table 3)
. At each site, the
SHOE condition significantly differed from
the other three treatments (collectively
termed healing footwear), in that force values when wearing the SHOE were greatest
at the heel, 4th MTH and 2nd MTH and least
at the midfoot site (Table 3)
. The TCC/WH,
TCC/CB and SPLINT, therefore, all significantly reduced force at these weight-bearing
locations of the foot and conversely increased force in the midfoot when compared
to the SHOE. The percentage change of
force with the healing footwear at each site
when compared to the SHOE is presented in
Figure 4
with the accompanying letter designation from the Duncan's test.
At the heel and 4th MTH, the significant
difference between the SHOE and the other
three treatment conditions was the only finding (Table 3)
. Additional differences were
found between the TCC/WH and the
SPLINT at the midfoot, and between the
TCC/WH and both the SPLINT and TCC/
CB at the 2nd MTH. In the midfoot, the
SPLINT resulted in the maximal force and
correspondingly the greatest increase when
compared to the SHOE (14.39 lbs., +/- 89.0
percent change). At the 2nd MTH, the TCC/
WH resulted in the least amount of, and
greater reduction in, force (14.80 lbs., -59.9
percent change).
Discussion
The fundamental benefit demonstrated by
use of the healing footwear was the reduction of force at the primary weight-bearing
areas of the foot, which correspond to the
most common sites of plantar ulceration and
is the basis for the clinical effectiveness in the
healing of plantar lesions (23). Additional
reduction of force is obtained clinically by
the addition of the Sifoam pad over the lesion, acting as an interface between the lesion and the plaster shell to remove all hard
weight-bearing surface away from the
wound. Figure 4
shows a minimum decrease
of -11.6 percent at the heel, -35.9 percent at
the 4th MTH, and -45.9 percent at the 2nd
MTH for all of the healing footwear conditions. This was achieved by providing total
contact to the entire plantar surface of the
foot, thereby creating a larger surface onto
which weight-bearing forces were distributed. Since weight-bearing forces, or body
weight, remain constant, the amount of
force applied to any specific unit area on the
plantar surface of the foot will be dependent
upon the total area available for distribution
of the total force. Increasing the total surface
area decreases the force exerted on a specified unit area of the foot, which is the concept of pressure, since pressure is defined as
the force per unit area.
The healing footwear effectively redistributes force and is demonstrated by the increases of +52.5 percent, +/-69.7 percent,
and + 89.0 percent in the midfoot for the
TCC/WH, TCC/CB and SPLINT conditions, respectively, when compared to the
SHOE (Figure 4)
. This area of the foot not
normally subjected to large weight-bearing
loads is now being used. Additionally,
weight-bearing forces are distributed to the
lower leg through the proximal extension of
the plaster to further unload the forefoot and
rearfoot.
To demonstrate the distribution of the
weight-bearing load within the plaster devices, Table 4 shows the force value at each
location, the total summed force and the percentage of the total summed force at all four
sites for that specific treatment condition.
Overall, the summed total decreased from a
value of 87.2 lbs. for the SHOE condition to
59.1 lbs. for the TCC/WH, 61.6 for the TCC/
CB, and 65.5 for the SPLINT, indicating a
reduction in applied force to the foot when
the healing footwear is worn. For the SHOE
condition, the 2nd and 4th MTH sites are
subjected to 37.2 percent and 24.3 percent of
the summed load, respectively. These values
are reduced to a maximum of 27.7 percent
for the 2nd MTH and 21.1 percent for the 4th
MTH with the healing footwear, showing a
shift from the forefoot to the rearfoot in the
magnitude of the percentage of the summed
weight-bearing load. Conversely, the percentage of the summed load at the midfoot
for the SHOE was 10.3 percent, which increased to a minimum of 20.4 percent for the
healing footwear. Surprisingly, although the
actual force value at the heel was reduced for
all the healing footwear conditions compared to the SHOE, the percentage of the
summed load borne by the heel actually increased from 28.2 percent in the SHOE condition to within a range of 30.3 percent to
35.6 percent for the healing footwear. The
following is a discussion of the results at each
of the four transducer sites.
Heel
The three healing footwear conditions
showed significantly lower forces than the
SHOE. The mean force values of 21.60 lbs.,
20.92 lbs., and 18.69 lbs. for the TCC/WH,
SPLINT, and TCC/CB conditions, respectively, translate to a - 11.6 percent, -11.9
percent and -20.6 percent decrease in force
when compared to the 24.60 lbs. recorded
for the SHOE. The decrease for the TCC/
CB condition (-20.6 percent) was nearly
double that for the TCC/WH and SPLINT
and could be due to the cushioning properties of the crepe sole. The degree of force
reduction at the heel was less than at the
forefoot sites and is consistent with existing
literature (20).
Midfoot
All three healing footwear conditions
showed a significant increase in force when
compared to the SHOE. This is due to better
utilization of the medial longitudinal arch as
a weight-bearing structure. The total contact
construction of the healing footwear raises
the insole of the device upward against the
arch to provide a base of support and allows
better contact by which to absorb weightbearing forces. By contrast, the relatively
flat insole of the shoe allows the arch to
maintain its more elevated position, thereby
reducing the load at the midfoot. This is similar in function to the localized relief built
into the healing footwear with the placement
of Sifoam padding over the lesion. In both
situations, the gap created between the plantar aspect of the foot and the underlying
footwear serves to minimize contact and reduce weight-bearing forces.
Additionally, differences were found
within the healing footwear group. The
SPLINT transmitted the greatest amount of
force to the midfoot (14.39 lbs.), and was
statistically different than the TCC/WH
(12.06 lbs.). These values, when compared
to the SHOE (8.98 lbs.), correspond to increases of + 89.0 percent for the SPLINT,
+ 69.7 percent for the TCC/CB, and + 52.5
percent for the TCC/WH.
Construction of the SPLINT and that of
the TCC/WH share some basic features,
such as the total contact design and the rubber walking heel. When comparing the results between these two devices at the midfoot, the differences can perhaps be explained by how well each is secured to the
leg. With the TCC/WH, the rigid plaster
completely surrounds the lower leg, reducing extraneous motion between the leg and
cast to a minimum. However, the SPLINT is
essentially a posterior splint and must be secured to the leg with an elastic wrap. This
provides less constraint on both vertical and
horizontal motion of the leg inside the
SPLINT and could permit greater compressive plantar forces. Shearing forces on the
plantar foot would also likely increase, although the system used in this study cannot
measure these forces. It must be noted, however, that despite the presumed increased
mobility and the measurable differences in
the midfoot and 2nd MTH (see below), healing rates for plantar lesions using either the
SPLINT or TCC/WH are comparable (5).
4th MTH
The only significant finding was the reduction in force of the healing footwear in comparison with the SHOE. The mean force values for the SPLINT (13.02 lbs.), TCC/CB
(13.01 lbs.), and TCC/WH (11.21 lbs.) represent reductions of -35.9 percent, -36.4 percent and -44.8 percent, respectively, when
compared to the SHOE (21.20 lbs.), and
they are more than twice the magnitude of
those for the heel. This greater reduction can
be attributed to proportionately larger percentages of the summed load being borne by
the heel and midfoot (Table 4)
when compared to the maximal load under the forefoot.
2nd MTH
The three healing footwear conditions all
showed significant reductions of force when
compared to the SHOE. The force values
17.16 lbs., 17.05 lbs., and 14.80 lbs. for the
SPLINT, TCC/CB, and TCC/WH, respectively, represent reductions of -45.9 percent, -46.8 percent, and -52.9 percent
when compared to the SHOE. The order of
decrease is similar to that for the 4th MTH,
with the SPLINT showing the least force
reduction, followed by the TCC/CB and
then the TCC/WH. The lower decrease with
the SPLINT can again be explained by the
greater degree of movement occurring within the SPLINT compared to the TCC/CB
and TCC/WH conditions. The magnitude of
reduction under the 2nd MTH ranges between 1.18 and 1.28 times the reduction
measured under the 4th MTH for each device.
Additional differences were found between the TCC/WH and the remaining two
devices. The 14.80-lb force value for the
TCC/WH is significantly less than the values
measured for the SPLINT and TCC/CB, and
the -52.9 percent force reduction is over 13
percent greater than the force reductions for
the SPLINT and TCC/CB. The difference
between the TCC/WH and TCC/CB conditions may be due to the rocker design of the
TCC/CB, which would allow more direct
force application under the forefoot during
the latter stages of stance phase as weight is
transferred forward over the front of the
TCC/CB. The lack of a difference between
the SPLINT and TCC/CB could again be
explained by movement within the SPLINT,
negating the benefit of the rubber walking
heel used with the SPLINT.
Summary
When compared to a conventional leather
oxford shoe, all three healing footwear devices were effective in reducing force at the
primary weight-bearing areas of the foot,
while increasing the distribution of force in
the midfoot region. The total contact design
of the devices more evenly distributed
weight-bearing forces throughout the entire
plantar surface of the foot and lower leg,
including those areas that do not normally
bear a large weight-bearing load. This reduces the amount of force applied to a specific unit area of the foot, which is equivalent
to reducing the pressure. The use of foam
relief over a lesion site during clinical application would further enhance the force reducing effects demonstrated in this study.
Referring to the three original purposes of
the study, the following conclusions are presented:
- a.) The SPLINT significantly reduced
the force in both the rearfoot and the
forefoot when compared to the
SHOE.
b.) The SPLINT resulted in significantly
greater forces distributed to the midfoot than the SHOE (and the TCC/
WH).
- a.) There was no difference between the
SPLINT and TCC/CB at the 2nd
MTH, but both were significantly less
effective than the TCC/WH.
b.) There were no significant differences
in force reduction between the
SPLINT and either the TCC/CB or
the TCC/WH at the heel and 4th
MTH.
- The only difference between the
TCC/WH and TCC/CB conditions
occurred at the 2nd MTH, with the
TCC/WH more effectively reducing
force at this site.
When considering the effect of the healing
devices on the different regions of the foot,
these conclusions can be made:
- The most effective device for reducing
force under the forefoot is the TCC/
WH, although the reduction is statistically significant only at the 2nd MTH.
- Under the heel, no statistically significant differences were found.
- At the midfoot, the SPLINT and TCC/
CB produced the greatest increase in
force measures, with the SPLINT being
most effective.
These findings provide quantitative data
illustrating significant force reduction of the
three devices compared to the SHOE to support their use in the healing of plantar lesions
(5). Although some statistical differences
were found among the healing footwear, the
magnitude of difference between the force
values was not very large (less than 2.4 lbs.).
Thus, regardless of the type of healing device
used (a walking splint or one of the two types
of total contact casts), or the type of terminal
device on the bottom of the healing footwear
(rubber walking heel or cast boot), similar
results were obtained.
This information will benefit all clinicians
involved in the management of pressure-related neuropathic ulcers by showing the effectiveness of these orthotic devices in reducing plantar pressure.
Andrew Novick, MA., P.T., is a research
physical therapist, Paul W. Brand Biomechanics
Lab, Rehabilitation Research, Gillis W. Long
Hansen's Disease Center, Carville, La. 70721.
James A. Birke, M.S., PT., is chief, Physical
Therapy Department, Gillis W. Long Hansen's
Disease Center, Carville, La. 70721.
Sandra L. Graham, MS., P.T., is senior staff
physical therapist, U.S. Public Health Service
Alaska Native Hospital, Anchorage, Alaska
95510.
Edward Koziatek, P.T., is chief, Physical Therapy Department, U.S. Coast Guard Academy,
New London, Conn. 06320.
References:
- Coleman WC, Brand PW, Birke JA. The
total contact cast. Journal of the American Podiatry Association 1984;74: 11:548-52.
- Brindley GW, Cofield RH. Local treatment
for neuropathic foot ulceration and osteoarthropathy. Foot and Ankle 1985;5:5:245-50.
- Pollard JP, LeQuesne LP. Method of healing diabetic forefoot ulcers. British Medical Journal 1983;286:436-7.
- Pring DJ, Casiebanca N. Simple plantar ulcers treated by below-knee plaster and moulded
double-rocker plaster shoe: a comparative study.
Leprosy Review 1982;52:261-4.
- Birke JA, Coleman WC, Koziatek E, Patout
CA, Novick A. Healing rates of plantar ulcers in
patients with diabetes mellitus and leprosy. Unpublished study 1990.
- Birke JA, Graham SL, Novick A, Coleman
WC. Methods of treating plantar ulcers. Physical
Therapy, February 1991;71:2:116-22.
- Helm PA, Walker SC, Pullium G. Total contact casting in diabetic patients with neuropathic
foot ulcerations. Archives of Physical Medicine
and Rehabilitation 1984;65:691-3.
- Kaplan M, Gelber RH. Care of plantar ulcerations: comparing applications, materials and
non-casting. Leprosy Review 1988;59:59-66.
- Sinacore DR, Mueller MJ, Diamond JE,
Blair VP, Drury D, Rose S. Diabetic plantar ulcers treated by total contact casting. Physical
Therapy 1987;67:10:1543-9.
- Walker SC, Helm PA, Pullium G. Total contact casting and chronic diabetic neuropathic foot
ulcerations: healing rates by wound location. Archives of Physical Medicine and Rehabilitation
1987 ;68:217-21.
- Diamond JE, Sinacore DR, Mueller Mi.
Molded double-rocker plaster shoe for healing a
diabetic plantar ulcer. Physical Therapy
1987;67: 10:1550-2.
- Joseph B, Joshua 5, Fritschi EP. The moulded double-rocker plaster shoe in the field treatment of plantar ulcer. Leprosy Review
1983 ;54:39-44.
- Gordon V. Treatment of the patient with
calcaneal and plantar pressure ulcers of neurogenic etiology. Journal of the American Osteopathic Association 1976;75:525-6.
- Gristina AG, Nicastro JF, Clippinger F, Rovere GD. Neuropathic foot and ankle patellartendon-bearing orthosis as an adjunct to patient
management, Orthopedic Review 1977;6:5 :53-9.
- Brand PW. The insensitive foot (including
leprosy). In: Jahass MH, ed. Disorders of the foot
1982;2: 1266-86.
- Brand PW. The diabetic foot. In: Ellenberg
M, Rifkin H, eds. Diabetes mellitus: theory and
practice 3rd Ed. 1983:829-49.
- Enna CD, Brand PW, Reed JK, Welch D.
The orthotic care of the denervated foot in Hansen's Disease. Orthotics and Prosthetics
1976;30: 1:33-9.
- Geary NPJ, Klenerman L. The insensitive
foot, Northwick Park experience. Leprosy Review 1987;58:79-84.
- McFarlane RM. The treatment of ulcers of
the heel. Geriatrics, April, 1965:326-31.
- Birke JA, Sims DS, Buford WL. Walking
casts: effect on plantar pressures. Journal of Rehabilitation Research and Development
1985 ;22:3: 18-22.
- Pollard JP, LeQuesne LP, Tappin JW.
Forces under the foot. Journal of Biomedical Engineering 1983;5:37-40.
- Birke JA, Nawoczenski DA. Orthopedic
walkers: effect on plantar pressures. Clinical
Prosthetics and Orthotics 1988;12:2:74-80.
- Birke JA, Sims DS. Plantar sensory threshold in the ulcerative foot. Leprosy Review
1986 ;57:261.
|
|